Installing the bellows

Hold the breathing system from above to prevent it
dropping out and
1 uniformly slide the tightly compressed bellows onto
the connector.
2 Insert the container in the breathing system from
below and turn it clockwise as far as possible.

Filling and installing the absorber

3 Push the insert fully into the absorber.
Fill the absorber with fresh soda lime
– up to the MAX mark.
Only use Drägersorb 800!
If the breathing system is to remain unused during the
next 24 hours:
Only fill with soda lime immediately before use!
Soda lime dries out when exposed to ambient air
and sun for a longer period of time. This reduces
the CO
absorption capacity and the effectiveness
2
of the indicator.
4 Insert the absorber in the breathing system from
below and turn clockwise as far as possible.
5 Slot in the breathing system – as far as possible.
6 Fold down the writing top.
1
2
MAX
3
6
5
5
Care
Assembling
4