Monitoring

Selecting the trend screen

Selecting the trend screen
Displays the measured values over time since the start of
measurement.
Maximum storable time: 8 hours.
The following display combinations can be selected:
– CO
/MV
2
– AGas/N
O
2
– O
/compliance*
2
– SpO
/pulse rate
2
S
Press the
key several times until the trend screen
appears.
AMV and compliance trends are scaled in accordance
with the setting in the configuration menu for adults or
children.
AMV trend scale:
0 to 3 L/min for children
0 to 15 L/min for adults
Compliance trend scale:
0 to 20 mL/mbar for children
0 to 100 mL/mbar for adults
Display (example):
Trend CO
and MV
2
Selecting other display combinations
Press the appropriate softkey:
»AGas N
2
or
»O
compl.«
2
or
»SpO
pulse«
2
The softkey disappears if a measuring function is not
available.
____________
*
Lung properties can be monitored continuously with the compliance
trend display so that ventilation can be adjusted accordingly.
Patient compliance is not measured in MAN/SPONT ventilation
mode.
58
IPPV
CO
PAW
2
60
40
30
0
11:00
12:00
20
MV
15
10
5
0
0
Select zoom area: Turn
!
Freshgas O
+ N
O
2
2
P
V
max
O
%
L/min
mbar
mL
2
50
1.00
25
600 10
98
67
alarm
SpO
2
limits
CO
CO
2
38
MV
_
etCO
freq
2
AGas
Ve
6.0
10
MV
N
13:00
O
Fi
Fet
compl.
40
36
SpO
O
2
pulse
0.8
0.6
Iso.
full
trend
58
56
N
O
2
Zoom in: Press
!
Freq.
T
:T
T
:T
PEEP
T
I
E
IP
I
1/min
%
mbar
1:2 10
2
O
2
2
2
0

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents