Installing The Breathing Hoses; Preparing The Suction System - Dräger Julian Instructions For Use Manual

Anaesthetic workstation
Hide thumbs Also See for Julian:
Table of Contents

Advertisement

Care
Assembling

Installing the breathing hoses

1 Fit the microbial filters on the inspiration and
expiration ports of the breathing system until they
audibly engage.
2 Connect the 2.3 L breathing bag to the breathing
hose with a connecting sleeve. Plug the breathing
hose onto the angled connector and hang the
breathing bag on the hook.
3 Plug the breathing hoses into the microbial filters
and connect them to the Y-piece.
Do not use antistatic or conductive breathing
hoses!
Fire hazard when using HF electrosurgery.
4 Fit the water trap container from underneath.
5 Screw the sample line onto the Y-piece and water
trap.
Only use original sample line. Other lines may
change the technical data of the device.
Alcohol and agents containing alcohol must not be
allowed to penetrate the sample line!
Alcohol distorts the results of the concentration
measurement.

Preparing the suction system

6 Plug the bacterial filter into the port on the back panel
of the Julian.
Replace the bacterial filter every two weeks.
The filter must be disposed of as infectious special
waste. It can be incinerated with little pollution at
temperatures of more than 800 °C.
108
1
3
2
5
4
6
5