Starting Up
Checking the workstation against the checklist
AGS anaesthetic gas scavenging system
1 The transfer hose from the waste gas port must be
connected.
2 The scavenging hose must be connected;
the anaesthetic waste gas probe must be plugged
into the Dräger wall socket and the indicator must be
green.
3 The float must be between the two marks.
Breathing system
Complete and engaged, with both valve discs
inserted, bellows fitted and hoses securely
connected.
Soda lime renewed, no violet discoloration.