Operation
Using non-rebreathing systems

In the event of a power failure

Switch off the external fresh gas output:
Press the softkey »Fresh gas internal/external« and
confirm by pressing the rotary knob.
Julian switches over to MAN/SPONT ventilation mode
or
Press the »IPPV« or »PCV« key and confirm with
the rotary knob.
Julian then switches over directly to controlled
ventilation via the original rebreathing system.
When changing from the external non-rebreathing system
to the rebreathing system of the Julian:
Reconnect the sample line to the Y-piece.
In the event of a power failure
Julian automatically switches over to the integrated
uninterruptible power supply (UPS).
If the battery is charged, operation will be maintained for
an average of 30 minutes.
The message
POWER FAIL!
appears on the screen
together with the remaining available battery capacity in
percent.
Example:
J J J J
95 %
If the battery is almost completely discharged, the
message BATTERY LOW !!! appears on the screen.
In the event of a power failure if the battery is discharged,
Julian switches over automatically to MAN/SPONT
ventilation mode. The fresh gas mixer is deactivated.
Version without rotary knob for O
metering:
Julian automatically delivers an O
This O
flow bypasses the Vapor unit, and so the patient
2
no longer receives any anaesthetic agent.
Version with rotary knob for O
Set rotary knob »Safety-O
metering to the desired O
Range 0 to 12 L/min.
This O
flow flows through the Vapor unit.
2
Check Vapor unit setting.
Ventilate patient manually!
52
emergency
2
flow of 4 L/min.
2
emergency metering:
2
« for O
emergency
2
2
flow.
2
IPPV
Pleth
CO
2
40
0
PAW
20
0
Freshgas O
+N
O
2
2
P
V
max
mbar
mL
O
%
L/min
2
50
1.00
25
600
D
POWER FAIL !
98
67
SpO
2
auto set
CO
2
38
limits
_
etCO
freq
2
Ve
6.0
10
alarm
MV
info
list
Fi
Fet
40
36
O
2
0.8
0.6
Hal.
config.
58
56
N
O
2
Freq.
T
:T
T
:T
PEEP
T
I
E
IP
I
1/min
mbar
%
10
1:2
10
Julian
J
95 %
alarm
limits
curves
0

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents