PCV ventilation mode

PCV = Pressure Controlled Ventilation
In pressure controlled ventilation mode, the applied tidal
volume depends on the ventilation parameters Pmax,
insp. Flow, insp. Time, PEEP and the lung compliance.
Changes in ventilation parameters or lung compliance
influence the tidal volume and the MV minute volume
must therefore be monitored constantly.
Presetting ventilation parameters for PCV
1 Press »PCV« its LED flashes.
The five ventilation parameters for PCV appear on the
screen with the default settings.
2 Press the softkey of the relevant ventilation parameter.
3 Set and confirm the ventilation parameter with the
rotary knob.
Display (example):
Adjustment ranges and default settings on delivery
Ventilation parameter
Pressure limit
Pmax [mbar]
Frequency Freq.
[1/min]
Insp./Exp. ratio
2: 1, 1: 1, 1: 2,
[T
: T
]
I
E
Insp. flow [L/min]
PEEP [mbar]
When ventilating infants weighing less than 20 kg:
Use the paediatric hose set;
see page 49 "Ventilating children".
–––––––––––
*
The default settings can be set specifically for each hospital,
see page 82.
** Can be set by DrägerService.
Adjustment
Default setting
range
on delivery*
(PEEP+1)
15
to 70
6 to 60
12
1 : 2
1: 3, 1: 4
5 to 50 (75**)
30
0 to 20
0
2
Standby
Julian
SW version 2.0
01. Nov. 98
To Start PCV : Confirm
Freshgas O
+N
O
2
2
P
max
O
%
L/min
mbar
2
15
30
4.00
Operation
Selecting the ventilation mode
Julian
3
1
1
leakage
test
delete
trend
config.
!
Freq.
T
:T
Insp.Flow
PEEP
I
E
1/min
L/min
mbar
12 1:2 30
0
PCV
43