Transmitting 12-Lead Patient Records; Troubleshooting - ZOLL E Series Operator's Manual

Defibrillator
Hide thumbs Also See for E Series:
Table of Contents

Advertisement

3. To deactivate automatic data transmission, display
the Summary menu (by pressing the SUMMARY
button) and press the Data Relay Off softkey
Note: If you change the destination port while a record
is already being transmitted, that transmission
continues to the previously selected port. All other
records in the queue will be transmitted to the
newly-selected port.
Note: If you change ports from Bluetooth to RS-232,
and then you want to change back to Bluetooth,
the host system must reestablish a connection
with the E Series unit. However, power to the
Bluetooth module is not turned on again until a
record starts transmitting out the Bluetooth port.

Transmitting 12-Lead Patient Records

You can transmit 12-lead ECG patient records to a
remote host system running ZOLL Data Relay software
without placing the E Series unit into Realtime mode.
1. With the E Series unit in Monitor mode (ON for AED
units), press the 12 Lead softkey to access the 12
Lead menu.
120
NIBP
SpO2%
24
80
98
100
38
CO2
m
24
Hg
RR
m
ECG
12 LEAD MONITOR
01:20
Acquire
Settings
PT Info
2. Press the PT Info softkey to access the Patient
Information menu.
120
NIBP
SpO2%
24
80
98
100
38
CO2
m
24
Hg
RR
m
PATIENT INFORMATION
Press Return to Exit to
12 Lead Monitor Mode
01:22
60
Male
ID #
Age
Gender
9650-1210-01 Rev. T
II
x2
120
ECG
Alar ms
Return
II
x2
120
ECG
:
Patient
Records
Return
3. Press the Patient Records softkey to display the
Patient Records menu.
17 JUN 04
17 JUN 05
14:07:32
13:24:03
17 JUN 05
11:52:48
17 JUN 05
Print
Print
Record
All
4. Select a Patient record and press the Transmit
softkey.
5. Select the Data Relay option and press the Dial
Phone# softkey.
If you are transmitting via Bluetooth, then the green
power LED on the E Series unit illuminates. You must
now establish a connection between the host system
and the E Series unit.

Troubleshooting

If configured, the unit displays the following messages
when it encounters problems during data transmission.
Message
Possible Causes and
Corrective Action
CARD FULL
Data card is full. No more data
can be stored on the card, but the
unit continues to operate. This
prompt is only issued when the
unit is not analyzing or charging.
Insert another card, or print event
data on stripchart recorder.
INSERT CARD
Card is not installed or not seated
properly in the unit.
CARD NOT
An Erase Card operation was
ERASED
interrupted, or the card is
PROPERLY /
damaged. The unit operates, but
ERASE CARD
no data is being stored.
Erasing the card may allow return
to normal data storage operation.
CARD WRITE-
The card is write-protected. The
PROTECTED
unit operates, but no data is
being stored.
Adjust the write-protect tab on
the card to allow writing.
200506171407
14:07:32
200506171324
200506171152
B-5