10.0 ECG Lead-Off Detection
Equipment Needed
•
Power Supply 12V 10A,
•
Battery Current Test Fixture (P/N 9100-0690-TF) or AED PRO Disposable Battery (P/N
1008-002003-01),
•
Defibrillator Analyzer (e.g. QED-6 or Impulse 4000) or equivalent
AED PRO 3-Lead ECG Cable (PN 9500-001802) or (8000-0838).
•
Setup:
Insert the AED PRO 3-Lead ECG Cable into the connector on the AED Pro unit and connect
•
the leads to the Defibrillator Analyzer.
•
Set the Defibrillator Analyzer to NSR (normal sinus rhythm).
•
Place the Battery Current Test Fixture in the battery well.
•
Set the Power Supply to 12.00V.
Test
10.1
10.2
10.3
10.4
9650-0309-01 Rev. E
Action
Turn on the Power Supply.
Turn on the AED Pro unit.
Pull any of the ECG leads from
the Analyzer.
Press and hold the On/Off
button to power off the unit.
ZOLL AED Pro Service Manual
Expected Result(s)
Verify the unit issues a voice
and text prompt UNIT OK
before unit shuts off by itself.
Verify the rhythm is displayed
and the unit issues a voice and
text prompt ECG
MONITORING CABLE.
Verify that the signal disappears
and the unit issues a voice and
text prompt CHECK ECG
ELECTRODES
Unit powers off.
Test Equipment
,
Pass
Fail
1–15