10.0 ECG Lead-Off Detection

Equipment Needed
Power Supply 12V 10A,
Battery Current Test Fixture (P/N 9100-0690-TF) or AED PRO Disposable Battery (P/N
1008-002003-01),
Defibrillator Analyzer (e.g. QED-6 or Impulse 4000) or equivalent
AED PRO 3-Lead ECG Cable (PN 9500-001802) or (8000-0838).
Setup:
Insert the AED PRO 3-Lead ECG Cable into the connector on the AED Pro unit and connect
the leads to the Defibrillator Analyzer.
Set the Defibrillator Analyzer to NSR (normal sinus rhythm).
Place the Battery Current Test Fixture in the battery well.
Set the Power Supply to 12.00V.
Test
10.1
10.2
10.3
10.4
9650-0309-01 Rev. E
Action
Turn on the Power Supply.
Turn on the AED Pro unit.
Pull any of the ECG leads from
the Analyzer.
Press and hold the On/Off
button to power off the unit.
ZOLL AED Pro Service Manual
Expected Result(s)
Verify the unit issues a voice
and text prompt UNIT OK
before unit shuts off by itself.
Verify the rhythm is displayed
and the unit issues a voice and
text prompt ECG
MONITORING CABLE.
Verify that the signal disappears
and the unit issues a voice and
text prompt CHECK ECG
ELECTRODES
Unit powers off.
Test Equipment
,
Pass
Fail