You can then view four 12-Lead snapshot pages by pressing
sequentially. For example, the unit displays the following 12-Lead snapshot after the Analysis
Page 2:

Fault Conditions Affecting 12-Lead Interpretive Analysis

The Propaq MD unit will not perform Interpretive Analysis if it detects one of the following
fault conditions when attempting to acquire 12-Lead data:
A pacer signal is present.
A lead fault is detected in the ECG cable.
An invalid cable is in use.
If the Propaq MD unit detects one of these fault conditions, Analysis Page 1 indicates that there
is No data available for Interpretive Analysis and lists the fault condition; all measurements on
Analysis Page 2 appear as N/A.
Once you have corrected the fault condition, press
the 12-Lead Interpretive Analysis.
9650-0806-01 Rev. D
Propaq MD Operator's Guide
12-Lead ECG Monitoring Setup
to step through them
to confirm the correction and perform
12
14-11