3.11 Basic Initial Functional Test

CAUTION:
are present on the input and output during this test.
Refer to Figure 3-19 for the required functional test set up. Proceed as follows to perform a
basic function check of the power system:
1. Verify the correct AC line input rating on the nameplate of the MX unit(s) and make sure the
correct three-phase line voltage is wired to the input of the MX before applying input power.
2. Connect a suitable resistive or other type load to the output of the MX. The load resistance
value will depend on the voltage range you plan to check. Make sure the power resistor has
sufficient power dissipation capability - up to 15 KW for full load test on one phase of MX45-
3 or 10 KW for full load test on one phase of M30-3Pi - and that the load used does not
exceed the maximum power rating of the MX. For three phase configurations, this test can
be performed on one phase at a time if needed.
3. Connect an oscilloscope and DMM / voltmeter to the AC source output. Set both for AC
mode.
4. If the correct voltage is present, turn on the MX unit(s) by closing the On/Off circuit breaker
on the front panel. For multi-cabinet systems, turn on the auxiliary unit first and wait for
them to cycle on, then turn on the master unit.
5. If the MX has more than one available output voltage range, go to the PROGRAM 1 screen
and select the desired voltage range. The output mode can be set from the PROGRAM 2
screen (use the MORE soft key or press the PROGR function key again). Select AC mode.
6. Set the output voltage to 0 volt and close the output relay with the OUTPUT ON/OFF button.
There should be little or no output although the DMM may show a noise level, especially if
the DMM is in auto ranging mode.
7. Move the cursor to the VOLTAGE field in the PROGRAM 1 screen and either use the
keyboard to program a small voltage (20 VAC) or slew the voltage up slowly with the knob.
Observe the DMM reading. The reading should track the programmed voltage.
8. Also monitor the scope. The output should be a sinusoidal voltage waveform.
9. If the output tracks, increase the voltage till you reach 80 % of the voltage range or more.
Check the output voltage reading and waveform.
10. Select the MEASUREMENT 1 screen by pressing the MEAS button. The output voltage,
current and power will be displayed. For three phase configurations, use the PHASE button
to select the øABC display mode. This will show the voltage, current and power for all three
phases. If all phases ar
all three unless the volta
loaded, current and power will only be shown for the loaded phase.
68
Work carefully when performing these tests; hazardous voltages
e loaded equally, the same curren
ges are not programmed to the s