Figure 9-29:Iec 61000-4-13 Test Flowchart Class 3 - California Instruments MX30-1 User Manual

Mx series ac and dc power source series i / series ii
Table of Contents

Advertisement

User Manual - Rev AJ
Start - Class 3
De
termine appropriate test
configuration
Perform 8.2.1 Test
"Harmonic combin
Flatcurve and Overswing
Any functional
anomalies?
Yes
Yes
Test Completed
FAIL
Record Results
MX Series
ation"
No
Yes
Perform 8.2.4 Test
"Meister Curve"
Any functional
anomalies?
No
Perform 8.2.3 Test
Only table 1. Odd
non-multiple of
3 harmonics
Yes

Figure 9-29:IEC 61000-4-13 Test Flowchart Class 3

No
Any functional
anomalies?
Perform 8.2.3 Test
"Individual Harmonics/
Interharmonics"
No
Any functional
anomalies?
California Instruments
Perform 8.2.2 Test
"Frequency Sweep"
Yes
Any functional
anomalies?
No
Test Completed
PASS
Record Results
221