9.4.6.2 VOLTAGE VARIATION TESTS
From the IEC1000-4-11 Main Menu screen shown in Figure 9-14, scroll to the VOLTAGE
VARIATIONS entry using the up and down cursor keys. Press the ENTER key to select the
VOLTAGE VARIATIONS menu. The screen will appear as shown in F igure 9-16.
RUN AL
L
The RUN ALL selection will cause the following automated test sequence suggested by the
standard to be run:
Step
Type
1
VSweep
2
Hold
3
VSweep
4
VSweep
5
Hold
6
VSweep
7
VSweep
8
Hold
9
VSweep
10
VSweep
11
Hold
12
VSweep
13
VSweep
14
Hold
15
VSweep
16
VSweep
17
Hold
18
VSweep
The user can change the NOMINAL Ut voltage for this test. The RUN ALL Command line will
change to ABORT during the test. Selecting ABORT and pressing the ENTER key will terminate
the test at any time and the output voltage will return to the nominal value.
210

Figure 9-16: Voltage Variation Screen

Start V in % of
U
T
100
40
40
100
40
40
100
40
40
100
0
0
100
0
0
100
0
0

Table 9-20: Voltage Variations Test Performed During RUN ALL

Dwell time
End V in % of
U
T
2
40
1
40
2
100
2
40
1
40
2
100
2
40
1
40
2
100
2
0
1
0
2
100
2
0
1
0
2
100
2
0
1
0
2
100
California Instruments
Delay between
steps (s)
0
0
10
0
0
10
0
0
10
0
0
10
0
0
10
0
0
10
MX Series