9.4.7 Using the MXGui Windows Program for IEC 61000-4-11 Testing
The MXGui Windows control program will detect the presence of the –411 option on the MX AC
power source. Test reports can be generated at the end of a test for documentation purposes.
To s pp rt Editi
u
o
is re uir d. Olde
q
e
can e s lected
b
e
Generic tests files are di
Files applicable to Edition 2.0 have ED20 in their file name. Do not mix these files, as the data
setup will not be correct if you do. To load a test file, select the Mode (Dips or a ) and test
standard revision first, then use the File, Open menu to load the test parameters. Test
parameters can be a function of the EUT class. The different files provided with the program
cover the various EUT classes. The relevant EUT class 1, 2, 3 or X is listed in the file names.
The user must select the desired test type before executing the test. Since both test types
require a number of test parameters, the test sequence parameters must be entered in the data
entry grid or loaded from disk using the File, Open menu entry.
For complete details on how to use the MXGui –411 option test screen, refer to the on-line help
of the MXGui program available from the Help menu.
212

Figure 9-19: IEC 61000-4-11 GUI screen.

on 2.0 of the IEC 61000-4-11 test standard, version 1.20 or higher of the MXGui
r versions of the MXGui will only support Edition 1.0 tests. The desired Edition
by the user and provides the relevant data en
stributed with the MXGui program for both editions of the test standard.