Figure 9-12: Abnormal State; Table 9-14: Normal Voltage Maximum; Table 9-15: Voltage Surge - California Instruments MX30-1 User Manual

Mx series ac and dc power source series i / series ii
Table of Contents

Advertisement

User Manual - Rev AJ
VOLT MAX
This test will change the output voltage from 28V or 14V to 30.3V or 15.1V. The test will last for
30 minutes. The ← (backspace) will terminate the test at any time.
VOLT UNDER
This test applies to category Z and 28 volt category B equipment. The output voltage will drop to
10 volts and will ramp up at a rate of 0.15 volt/sec for the US standard and at a rate of 0.30
volt/sec for EUROCAE standard for 30 seconds before it returns to nominal value.
VOLT SURGE
This test will surge and sag the voltage to a
seconds between transients. The test is repeated three times.
C
ategory
A
B
Z
POWER INTERRUPT
Refer to section POWER INTERRUPT.
9.3.1.6.2 Abnormal Test
From the DO-160 MENU scroll to the ABNORMAL DC entry using the up and down cursor keys.
Press the EN
TER ke to select the ABNORMAL screen. The screen will app
Figure 9-12.
The Abnormal Test has the following tests:
1. VOLT MIN
2. VOLT MAX
3. VO
LT LOW
4. VOLT DROP
5. VOLT SURGE
MX Series
Standard
A and Z
RTCA
30.3
EUROCAE
30.3

Table 9-14: Normal Voltage Maximum

Su
rge
Volt
Dwell(msec)
RTCA
EUR
40
40
40
40
50
50

Table 9-15: Voltage Surge

y

Figure 9-12: Abnormal State

Categories
B 28V / 14V
30.3
29.3
level and dur
ation sp
Volt
U
S
EUR
30
1
5
17
30
15
17
5
0
12
12
California Instruments
15.1
14.6
ecified Tab
le 9-15
with
Sags
Dwell(msec)
30
30
30
ear as shown
5
203