4.7.4 Pulse Transients
Pulse transients let you program the output to a specified value for a predetermined am
time. At the end of the Pulse transient, the output voltage returns to its previous value.
Parameters required to set up a Pulse transient include the pulse count, pulse period, and pulse
duty cycle. An example of a Pulse transient is shown in Figure 4-42. In this case, the count
the pulse period is 16.6 ms or 60 Hz and the duty cycle is 33%.
Note that Pulse transients can only be programmed over the bus, not the front panel. Refer to
the SCPI Programming Manual for more information about programming Pulse transients and
triggers.
138

Figure 4-42: Pulse Transients

California Instruments
ount of
is 4,
MX Series