Figure 4-39: Set Volt Trigger Source Acquisition - California Instruments MX30-1 User Manual

Mx series ac and dc power source series i / series ii
Table of Contents

Advertisement

User Manual – Rev AJ
Voltage step (SET VOLT)
MX Series
When selecting this trigger source, the field below the TRIG
SOURCE field changed to "TRIG PHASE =". Use this field to
enter the desired voltage phase angle to trigger the
measurement on.
This mode is appropriate when capturing analyzing events at a
specific phase angle such as the zero crossing of the voltag
Note that the phase angle of the current with respect to the
voltage is determined by the load, so triggering at a specific
phase current angle is not possible as it is not controlled by t
AC source. However, when capturing current waveform data
the phase relationship to the voltage can be determined easily
by triggering at the 0° point on the voltage.
This mode performs two functions. It programs the output
voltage for the selected phase or phases to the rms or DC va
specified and it triggers the measurement acquisition at the
same moment in time.
When selecting this trigger source, the field below the TRIG
SOURCE field changed to "SET VOLT =". Use this field to enter
the desired voltage to program the output to and trigger the
measurement on. If only one phase in a thre
selected, only that phase's output will be program
phases are selected, all three phases' outputs w
programmed. Use the PHASE key to select the desired p
or all phases. Figure 4-39 sh
VOLT trigger source to capture the turn-on of the voltage. In
this case, a negative trigger delay was specified and the voltag
start phase angle
START
[ENTER]
TRIGGER
DELAY
TRIGGER =
SET VOLT 120

Figure 4-39: SET VOLT Trigger Source Acquisition

California Instruments
ows an example of using the SET
was set to 90° in the PROGRAM 2 screen.
ACQUISITION WINDOW
e.
he
,
lue
e phase system is
med. If all
ill be
hase
e
133