Figure 9-22: Iec 1000-4-13 Fcurve; Figure 9-23: Iec 1000-4-13 Oswing - California Instruments MX30-1 User Manual

Mx series ac and dc power source series i / series ii
Table of Contents

Advertisement

User Manual - Rev AJ
9.5.4.2 IEC GROUPS
This section will describe the groups and parameters associated with IEC 1000-4-13. Refer to
paragraph 9.5.4.1 for groups associated with the test.
FCURVE GROUP
If the FCURVE group is selected, the screen will appear as shown in Figure 9-22. The screen
has the following parameters that are unique to the group:
1. LEVEL
2. PAUSE
OSWING GROUP
If the OSWING group is selected, the screen will appear as shown in Figure 9-23. The screen
has the following parameters that are unique to the group:
1. LEVEL
2. HARM NO There are only two harmonic numbers. This field is used to set the harmonics
3. PHASE
4. PAUSE
216

Figure 9-22: IEC 1000-4-13 FCURVE

Set the Flat curve clip level. This field can be set when the IEC 1000-4-13 is in
the OFF state and the Class is set for USER class.
Time in seconds the test will pause prior to proceeding to the following test. This
field can be set at any time when the test is not running.

Figure 9-23: IEC 1000-4-13 OSWING

The level of the harmonics relative to the fundamental in percent. There are
two levels, one for harmonic #3 and the other for harmonic #5. The harmonic
number must be selected prior to the harmonic level entry. To change levels,
the test state must be off and USER class must be selected.
level that corresponds to the harmonics number.
Sets the phase angle of the harmonics relative to the fundamental. The phase
can be set when the test state is off and USER class is selected.
Sets the time in seconds. The test will pause prior to proceeding to the
following test. This field may be changed at any time as long as the test is not
running.
California Instruments
MX Series

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents