Table 9-17: Phase Mapping; Table 9-18: Iec 61000-3-34 Table C.2 - California Instruments MX30-1 User Manual

Mx series ac and dc power source series i / series ii
Table of Contents

Advertisement

User Manual - Rev AJ
MX units with revisions prior to this will change the L-N amplitude by the set dip level but retain
the nor
mally programmed phase angle relationship. For firmware upgrades, contact
support@calinst.com.
Note that required phase angles and amplitudes are automatically set for dips of 0, 40, 70, 80
and 100% to conform with method (A). For all other dip levels, method (A) can be used by
programming the required phase angles to be used during the programmed dips. The amplitude
and phase angles required to obtain the correct line-to-line voltage dip per method (A) for
s
tandard dip le
vels
IEC 61000-4-34.
S
ince all phas
e pro
phase angle for A other than 0° are implemented by offsetting all three phases by the required
number of degrees to get phase A at 0°. This is reflected in the actual output settings shown on
the MX versus the data in table C.2 of the IEC61000-4-34. The actual output settings are shown
in the last 3 columns.
Phase Mapping
The phase rotation on the MX series is ACB. This means phase A is mapped to L1, phase B is
mapped to L3 and phase C is mapped to L2. The required phase selection letter combination
for the required Line-to-line dip is shown in table C.2 for reference.
To select the desired phase-to-phase dip, select the phase selection as shown in column 8 and
either 80, 70 or 40 % dip level from the IEC411 screen or the Gui. Table C.2
Line to Line
L1-L2
L2-L3
100 %
100% dip
150°
(no dip)
80%
80% dip
150°
L1-L2
92%
80% dip
161°
L2-L3
100%
80% dip
150°
L1-L3
70%
70% dip
150°
L1-L2
89%
70% dip
167°
L2-L3
100%
70% dip
150°
L1-L3
40%
40% dip
150°
L1-L2
87%
40% dip
187°
L2-L3
100%
40% dip
150°
L1-L3
206
of 40, 70 and 80% are embedded in the firmware and conform to table C.2 of
gramming on the MX series is referenced to phase A, voltage dip with a
IEC Tables
L1
L2
L3

Table 9-17: Phase mapping

L3-L1
L1-N
100 %
100 %
100 %
270°
30°
100%
92%
72%
270°
41°
14°
80%
100%
100%
270°
30°
92%
80%
100%
281°
30°
100%
89%
61%
270°
47°
25°
70%
100%
100%
270°
30°
89%
70%
100%
287°
30°
100%
87%
53%
270°
67°
79°
40%
100%
100%
270°
30°
87%
40%
100%
307°
30°

Table 9-18: IEC 61000-3-34 Table C.2

MX Reference
Line to Neutral
Phase
L2-N
L3-N
Selection
100 %
100 %
120°
240°
100%
100%
120°
240°
72%
230
134°
240°
100%
72%
120°
254°
100%
100%
120°
240°
61%
100%
145°
240°
100%
61%
120°
265°
100%
100%
120°
240°
53%
100%
199°
240°
100%
53%
120°
319°
California Instruments
A
C
B
MX Setting
A-N
C-N
n/a
100 %
100 %
120°
AC
72%
100%
106°
BC
100%
72%
134°
AB
100%
100%
120°
AC
61%
100%
95°
BC
100%
61%
145°
AB
100%
100%
120°
AC
53%
100%
41°
BC
100%
53%
199°
AB
100%
100%
120°
B-N
100 %
240°
100%
226°
100%
240°
72%
254°
100%
215°
100%
240°
61%
265°
100%
161°
100%
240°
53%
319°
MX Series