9.5 ENVIRONMENTAL SPECIFICATIONS

Reference Temperature:
Operating Temperature:
Storage Temperature:
With batteries:
Without batteries:
Recharging Temperature:
Altitude
Operating: 0 to 2000 meters (6560 ft)
Non-Operating: 0 to 10,000 meters (32800 ft)
Pollution degree 2
Indoor Use
WARNING: Above 104°F (40°C) the instrument must be powered by the battery alone OR from AC power with the
battery removed. Use of the instrument with both the battery AND AC power is prohibited.

9.6 SAFETY SPECIFICATIONS

Electrical Safety
IEC 61010-1; 600V CAT IV or 1000V CAT III
● The use of the AmpFlex
● The use of the MR193, MN93, MN193, J93 and SL261 probes rate the "device + current probe" at 300V CAT IV
or 600V CAT III.
● The use of the 5A adapter rates the "device + current probe" at 150V CAT IV or 300V CAT III.
Electromagnetic Compatibility
Emissions and immunity in an industrial setting compliant with IEC 61326-1. For electromagnetic emissions, the
instrument belongs to group 1, class A, under standard EN55011. Class A devices are intended for use in industrial
environments. There may be difficulties ensuring electromagnetic compatibility in other environments, due to conducted
and radiated disturbance.
For immunity to radio frequency fields, the instrument is intended for use at industrial sites under standard IEC61326-1.
For the AmpFlex
and MiniFlex
®
● An (absolute) influence of 2% may be observed on the current THD measurement in the presence of a radiated
electric field.
● An influence of 0.5A may be observed on the RMS current measurement in the presence of conducted radio
frequencies.
● An influence of 1A may be observed on the RMS current measurement in the presence of a magnetic field.
Specifications are subject to change without notice.
Power Quality Analyzer PowerPad
20 to 26°C (68 to 78.8°F) from 45 to 75% RH
0 to 42.5°C (32 to 108.5°F) from 10 to 85% RH
0 to 50°C (32 to 122°F) from 45 to 75% RH
-20 to 35°C (-4 to 95°F) from 0 to 95% RH
-20 to 50°C (-4 to 122°F) from 0 to 75% RH
-20 to 70°C (-4 to 158°F) from 0 to 75% RH
0 to 40°C (32 to 104°F) max
and SR probes rate the "device + probe" at 600V CAT IV or 1000V CAT III.
, MiniFlex
®
®
sensors:
®
III Model 8336
®
131