4.1.5 PROBES AND RATIOS

4.1.5.1 CURRENT PROBES

The
screen defines the current probes and ratios.
This menu enables you to define
■ Current probes (press A)
■ Current ratios (press V or U, depending on set-up)
Press A to define the current probes and ratios. This displays the current probe models detected by the instrument. If a
probe is detected but not managed, an error message is displayed.
For a 3-wire 3-phase set-up where only two of the three current sensors required are connected, if these two sensors are
of the same type and have the same ratio, the instrument simulates the third sensor by assuming the same characteristics
as for the two others. The third sensor will be shown in the list as either being simulated, or impossible to simulate.
The current probe choices that are available are:
If an MN193 clamp (5A range) or an adapter is used, the current ratio setting is proposed automatically. If an MN193
clamp (5A range), an Adapter, an AmpFlex
or sensitivity is proposed automatically.
Use the arrow buttons to define the ratios between the primary current (1A to 60,000A) and the secondary current (1A,
2A or 5A), then confirm the selections with the
current.
Power Quality Analyzer PowerPad
MN93 clamp: 200A
MN193 clamp: 100 or 5A
SR193 clamp: 1000A
J93 clamp: 3500A
AmpFlex
MiniFlex
MR193 clamp: 1000A/1200A
SL261 clamp: 100A (sensitivity 10mV/A
SL261 clamp: 10A (sensitivity 100mV/A
Three phase adapter: 5A
, a MiniFlex
®
button. Note that the primary current cannot be less than the secondary
III Model 8336
®
Figure 4-13
AC
AC
AC
/5000A
AC
DC
®
A193: 100A/6500A/10,000A
MA193: 100A/6500
®
AC
AC/DC
AC
, J93 or an SL261 clamp is used; adjustment of the ratio, range,
®
AC
)
AC/DC
)
AC/DC
29