Item
Function
Transient name, number, date, time and triggering channel of transient.
1
Activates/deactivates the choice of a transient list display filter.
2
The displayed value indicates the page number and the number of pages.
3
These icons let you browse through the previous and next screens. Press the yellow buttons corresponding to these icons to display
4
the pages.
Transient display filter:
- : displays the 4 single voltages during the transient (here in a 5-wire 3-phase connection)
- 4V: displays the 4 single voltages during the transient (here in a 5-wire 3-phase connection)
5
- 4A: displays the 3 currents and the neutral current during the transient
- L1, L2 or L3: displays the current and the single voltage successively in relation to phases 1, 2 and 3
- N: displays neutral
To select a transient, move the cursor to it using the ▲ and ▼ buttons. The selected field is bolded.
Validate with the
. button. The screen then displays the transients in graph form.
Power Quality Analyzer PowerPad
1
2
3
4
1
2
3
4
III Model 8336
®
Figure 5-4
Figure 5-5
5
5
6
43