Use the ▲ and ▼ buttons to choose sliding or non-sliding.
● Sliding: The PLT is calculated every 10 minutes. The first value is available 2 hours after the instrument is
switched on because it takes 12 values of PST to calculate the PLT.
● Non-sliding: The PLT is calculated every 2 hours.
Press
to confirm the selection and return to the Configuration menu.
NOTE: This setting cannot be modified when the instrument is recording, and/or searching for alarms.

4.1.4 ELECTRICAL HOOK-UP

The
menu is used to define how the Model 8336 is connected, according to the type of network.
Several electrical diagrams can be selected. Use the arrow buttons to choose a connection. One or more types of network
correspond to each distribution system.
These connection choices allow you to connect the instrument to all existing networks.
Press
to confirm the selection and return to the Configuration menu.
NOTE: This setting cannot be modified when the instrument is recording, and/or searching for alarms.
Distribution System
Single-phase 2-wire (L1 and N)
L1
N
Single-phase 3-wire (L1, N and ground)
L1
N
Power Quality Analyzer PowerPad
ELECTRICAL HOOK-UP
L1
L2
L3
N
Single-phase 2-wire non-grounded neutral
Single-phase 3-wire grounded neutral
III Model 8336
®
03/10/14 10:26
3-phase 5-wire
Figure 4-12
Source
am
3V
V1V2
V2V3
V3V1
L1
N
L1
N
GND
25