Environmental Specifications - AEMC SR661 User Manual

Ac current oscilloscope probe
Table of Contents

Advertisement

Drop Test: 1m (IEC 68-2-32)
Mechanical Shock: 100g (IEC 68-2-27)
Vibration: 5 to 15Hz, 0.15mm (IEC 68-2-6)
15 to 25Hz, 1mm
25 to 55Hz, 0.25mm
Output: Insulated lead with insulated BNC Connector, 6ft (2 m)
3.3

Environmental Specifications

Operating Temperature/RH: 14° to 122°F (-10° to 50°C)
Storage Temperature/RH: -4° to 158°F (-20 to 70°C)
Altitude: Non-operating: 0 to 12,000 m
Operating: 0 to 2000 m
Operating Relative Humidity: 0 to 85% @ 35°C
Influence of Temperature: ≤ 0.15% / 10K
Influence of Humidity: 10 to 90%: 0.1%
3.4
Safety Specifications
E192383
Electrical: Double insulation or reinforced insulation between the primary
or secondary and the outer case of the handle per EN 61010-2-032.
600V, CAT III, Pollution Degree: 2
Dielectric Strength: 5550V, 50/60Hz between primary, secondary, and
the outer case of the handle
Electromagnetic Compatibility: EN 50081-1 Class B
*Specifications are subject to change without notice
AC Current Oscilloscope Probe Model SR661
EN 50082-2 Electrostatic discharge
IEC 1000-4-2
Radiated field IEC 1000-4-3
Fast transients IEC 1000-4-4
Magnetic field at 50/60 Hz IEC 1000-4-8
9

Advertisement

Table of Contents
loading

This manual is also suitable for:

2113.49