Average Battery Life: The operating time will be 15 days or 3 weeks, based on
a 10-minute PI measurement.
Recharge Time:
6 hours for 100% capacity (10 hours if the battery is completely drained).
0.5 hours for 10% capacity (charge life: 2 days approximately).
NOTE: It is possible to recharge the batteries while performing
insulation measurements provided that the values measured are higher
than 20MΩ. In this case, the recharging time is higher than 6 hours and
depends on the frequency of the measurements.
7.5 ENVIRONMENTAL SPECIFICATIONS
Operating Range:
14° to 104°F (-10° to 40°C) during recharging of batteries.
14° to 131°F (-10° to 55°C) during measurement.
10 to 80% RH.
Storage: -40 to 158°F (-40 to 70°C); 10 to 90% RH.
Altitude: <2000m.
7.6 MECHANICAL SPECIFICATIONS
Case Dimensions:
10.63 x 9.84 x 7.09" (270 x 250 x 180mm).
Weight: 9.5 lbs (4.3kg) approx.
Mechanical Protection:
IP 53 per NF EN 60529 (Ed. 92).
IK 04 per NF EN 50102 (Ed. 95).
7.7 SAFETY SPECIFICATIONS
Electrical safety as per IEC/EN 61010-2-030 or BS EN 61010-2-030, IEC 61557.
1000V CAT III, Pollution Degree 2.
Electromagnetic Compatibility:
IEC/EN 61326-1.
44
Megohmmeter Models 5050 & 5060