Checking The Analyzer - Siemens ADVIA 1800 Operator's Manual

Chemistry system
Table of Contents

Advertisement

3. In the Mode set list, select
4. Verify the selections for the startup you want to run.
If no startup is defined or you want to change an existing one, proceed as follows:
a. For each of the Proc. set, Set1, Set2, or Set3, you can select one of the following:
b. Select
c. To confirm, select
5. In the Select wash routine area, select
6. In the Proc.set list, select the startup setting (Set1, Set2, or Set3) you want to run.
7. Select
8. To confirm, select
The current operating mode displays in the System mode box, and the time remaining to
complete the startup displays in the Time until end box. If a cell blank is run, the system
prompts you to save the results. When done, the system enters the Ready mode.
9. To halt the startup, do the following:
a. At the System Startup/Shutdown Settings window, select
b. In the Select wash routine area, select
c. At the Operation Panel, select

Checking the analyzer

1. Inspect the following components:
NOTE:
Section of this guide or to the online Help for step-by-step procedures.
• probes
• mixing rods
• dilution-cuvette washers (DWUD)
• reaction-cuvette washers (WUD)
• probe wash cups
• cuvette covers
• pumps for leaks
System Overview
Start set
(1) In the PRIME list, select the prime you want run or select
A Prime is required only if you replenish a system reagent or replace a
related component.
(2) Select the check box next to each wash you want to perform.
(3) Select
WASH3
for the startup wash.
(4) If you want to run Cell blank, select the check box next to Cell blank.
Run Cell blank weekly.
.
Save
.
Yes
.
Start
.
Yes
If any of the components require maintenance, refer to the Maintenance
.
Startup
.
.
Cancel
.
Stop
NONE
.
.
Stop
73