Checking and replenishing the lamp coolant

Materials required:
• Lamp coolant additive (REF
04533710, PN B01-4496-51)
• Deionized water
Time: 5 minutes
Analyzer mode: READY
The lamp is cooled by circulating liquid coolant. As the volume of coolant decreases, the
heat of the lamp increases.
NOTE:
coolant warning and turns off the lamp.
1. Lift off the lamp access cover (
Figure 5-13. Lamp access cover
2. Check the fluid level in the reservoir.
If the level is between the 5- and 9-cm marks, the label is green on the reservoir, proceed
to step 4.
132
49
in the CTT cup position 1st time field and
2nd time field.
in the RTT1 and RTT2 cup positions 1st time fields and
55
and RTT2 cup positions 2nd time fields.
.
Execute
Perform the lamp energy check and cuvette blank measurement only once a
Check the lamp coolant level daily and whenever the system generates a lamp
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
to gain access to the lamp coolant reservoir.