Siemens ADVIA 1800 Operator's Manual page 116

Chemistry system
Table of Contents

Advertisement

Figure 5-2. Location of DPP shield screw
3. Using a Phillips screwdriver, remove the screws that secure the DPP shield to the
analyzer panel. (See Figure 2.)
4. Push the DPP shield to the right and slowly lift the DPP shield until it reaches
approximately a 90° angle, then gently lift the tab of the DPP shield and remove.
5. Unscrew all thumb screws, then remove the splash guard protective covers from the
wash cups.
NOTE:
the splash guard protective cover.
6. Lift and manually rotate the probe arm to an accessible location.
The movement may feel a bit awkward and tight.
Figure 5-3. Manually adjusting probe
116
If the system includes an anti-rotation bracket, avoid hitting it while removing
CAUTION
Manually support (lift) the probe to avoid damaging the probe tip. Be careful
not to strike the probe against other components on the analyzer.
ADVIA 1800 Operator's Guide