Inspecting And Cleaning The Reaction (Wud) And Dilution (Dwud) Cuvette Washers - Siemens ADVIA 1800 Operator's Manual

Chemistry system
Table of Contents

Advertisement

6. Place a thermometer inside the water reagent wedge and measure the temperature. The
thermometer should read from 8° to 14°C.
NOTE:
provider.
7. Record the temperatures as part of the daily maintenance report.
Inspecting and cleaning the reaction (WUD) and dilution (DWUD)
cuvette washers
Materials required:
• Lint-free towel
• Alcohol prep pads or lint-free towels
and 5% bleach solution
• 4-mm hex wrench
Time: 10 minutes
Analyzer mode: STANDBY
1. Set the analyzer to Standby.
2. Inspect the exterior of the reaction cuvette washer (WUD) and dilution cuvette washer
(DWUD) tubing for cleanliness.
3. Check the WUD and DWUD for leaks.
TIP
Perform this inspection in addition to the startup, shutdown, and weekly washes, to
keep the WUD and DWUD from clogging.
In the event of a clog, call your local technical support provider or distributor for
assistance.
Figure 5-5. Cuvette wash stations
4. Remove the wash head:
a. Cover nearby cuvettes with lint-free toweling to protect them from dust.
b. Loosen the retaining screw (
c. Lift the wash head from the wash station assembly.
Maintenance
If the temperature reading is outside this range, please contact your service
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
1
) with a 4-mm hex wrench.
1
Dilution Cuvette Wash
Station (DWUD)
2
Reaction Cuvette Wash
Station (WUD)
121

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents