10. Return the analyzer to the READY mode.
11. Verify the wash head nozzles are correctly centered in the cuvettes:
a. At the Menu Panel, select
b. At he Manual Operation window, double-select
c. Select
d. Verify the nozzles are correctly centered.
e. At the DWUD or WUD window, select
12. At the Operation Panel, select
the up position and the instrument is in the READY state.

Inspecting and cleaning the cuvette splash covers

Materials required:
• Lint-free towel
• Deionized water
Time: 5 minutes
Analyzer mode: STANDBY
NOTE:
entering the dilution and reaction cuvettes
1. Inspect the cuvette covers for spills and splattering.
If there is any splattering on the cuvette covers (
Figure 5-7. Cuvette covers
Maintenance
For additional information concerning the Manual Operation window, refer to the
Manual Operation window in the online Operator's Guide.
to slightly lower the washer nozzles, then verify that they are
Move
correctly positioned.
If not, call your local technical support provider or distributor.
nozzles.
Cuvette covers are installed around the probes to prevent water and reagent from
, then select
Maint
Manual Operation
14.DWUD
Init
., then select
Initialize
, then verify the DWUD and the WUD are in
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
.
), proceed to step 2.
1
.
or
.
23.WUD
Exit
to raise the washer
123