Replacing The Reaction And Dilution Cuvettes - Siemens ADVIA 1800 Operator's Manual

Chemistry system
Table of Contents

Advertisement

4. Clean the old filter.
NOTE:
filters and place them on the analyzer to avoid downtime.
5. Place the filter in a beaker filled with a freshly made 10% solution of household
bleach and water.
6. After 30 minutes, remove the filter and rinse it with deionized water.
7. Reinstall the filter in the holder and screw the bottom half of the holder in place.
8. Using a pad soaked in alcohol, clean the outside surfaces of the filter holder and hose.
9. Insert the silicon filter hose into the water bottle.
10. Prime the lines:
a. Return the analyzer to the READY mode.
b. At the Operation Panel, select the
c. In the PRIME Set dialog box, select
d. Select

Replacing the reaction and dilution cuvettes

Materials required:
• 13 sets of reaction cuvettes (sample
cell RRV, single cuvette set,
REF 05024992, PN 073-0023-02)
• 6 sets of dilution cuvettes (sample cell
DTT, single cuvette set,
REF 05049669, PN 073-0022-01)
Time: 20 minutes
Analyzer mode: STANDBY
Replace the 13 sets of reaction (RRV) cuvettes and 6 sets of dilution (DTT) cuvettes
every four months.
Maintenance
An additional set of filters is included with the analyzer. Locate the spare
CAUTION
Ensure the filter is properly positioned within the filter holder to avoid it
shifting out of place. Store filters in a clean and dry container for later use.
number of times in all fields.
.
Execute
button.
PRIME
and type
PRIME 2
10
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
or more for the
155