Exporting The Sample Log Entries; Reviewing The Calibration Results; Reviewing The Sample Results; Reporting Results - Siemens ADVIA 1800 Operator's Manual

Chemistry system
Table of Contents

Advertisement

Exporting the sample log entries

You can export the current sample log file in ASCII format for use by another program
such as Microsoft Excel.
1. At the File menu, select
2. Select or create the folder in which you want to save the file.
3. In the File name box, type the file name you want.
The sample log file name is entered by default.
4. In the Save as type list, select the file format.
The CSV (comma separated values) file format is selected by default.
5. In the Deliminator list, select the character used to separate sample log data items.
6. Select

Reviewing the calibration results

Use the View Calibration Curve window to review the calibration details for
photometric methods.
Use the ISE Monitor to review the calibration details for electrolyte methods.
See Checking ISE Calibration for information about the criteria used to accept or reject the
ISE calibration.

Reviewing the sample results

Sample results can be reviewed on and report from the RealTime Monitor and
Review/Edit windows.

Reporting results

The system prints calibration, control, and patient sample results. Only patient and control
sample results are transmitted to a host computer.
Results are reported automatically as the sample analysis is complete and results become
available, or the operator can manually select results to report in batches.

Real-time reporting

During operation, the System set area of the System Monitor window manages the
printing of the sample results (usually for laboratory purposes).
The Automatic transfer area of the Online Settings window manages the real-time
transmission of the control and patient samples to a host computer.
Use the Data Clean feature of the Online Settings window to prevent the transmission of
questionable sample results. The Transfer Results List dialog box of the Review/Edit
window indicates which samples failed this data check and were not transmitted.
System Overview
Export
.
Save
.
93