error
G
Crash
h, l
Normal value
(upper,
lower)
H, L
Abnormal
value
(upper,
lower)
i
Ratio results
j
Exceeded the
calibration
high range
K
Maximum
absorbance
limit
exceeded
k
Exceeded the
calibration
low range
M
Mix error
n
Abnormal
number of
effective
points
N
Cell blank
P
Prozone
Q
Liquid level
sensor error
s
220
temperature is outside the
pre-defined range
A probe crash is detected
Result exceeds expected
range
Result exceeds expected
range
Ratio component test is
flagged
The calibration high range is
exceeded.
The maximum absorbance
limit is exceeded.
NOTE This flag is linked to
the Max Limit field at the
Analytical Parameters
Chemistry window.
The calibration low range is
exceeded.
A mixer motion error is
detected
Insufficient valid data points
Abnormal cell blank
Failed prozone check
A liquid level sensor error
occurs
Photometric test insufficient
sample
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
N
Y
Y
Y
Y
Y
Y
Y
ADVIA 1800 Operator's Guide
N
Y
Y
N
N
N
N
__
N