Cleaning And Replenishing The Cuvette Conditioner Bottle - Siemens ADVIA 1800 Operator's Manual

Chemistry system
Table of Contents

Advertisement

4. Immerse the cuvette segments in 5% Probe Wash 3 solution.
a. Ensure no air bubbles are in the cuvettes.
b. Allow the cuvettes to soak for at least 10 hours.
5. Wash the cuvettes under running water, then rinse them in deionized water.
6. Drain the water from the cuvettes.
7. Install the cuvette segments on the dilution tray (DTT) and fasten the thumbscrews
by hand.
8. Return the analyzer to the READY mode.
9. At the Operation Panel, select
NOTE:
further actions.
10. At the Operation Panel, select
11. Perform the daily shutdown wash routine, then verify the operation.

Cleaning and replenishing the cuvette conditioner bottle

Materials required:
• Deionized water
• Cuvette conditioner
Time: 10 minutes
Analyzer mode: STANDBY
NOTE:
cleaned at least once every 2 months.
WARNING
Figure 5-21. Cuvette conditioner bottle
Maintenance
Verify that the Operating mode field displays READY before performing any
The cuvette conditioner bottle may be cleaned when it is refilled, but must be
Before handling any reagents, read the warnings on page 196
Initialize
.
WASH
.
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
1
2
3
Cuvette Conditioner Bottle
Level Sensor Connector
Filter Cap
147