Additional Operating Instructions; Switching Process Sequence And System Test List Numbers; Loading The Reagent Trays; Updating The Cumulative Qc - Siemens ADVIA 1800 Operator's Manual

Chemistry system
Table of Contents

Advertisement

3. Wait approximately 20 seconds.
4. Perform the normal startup operation and open the Startup window.
5. At the Startup window, enter a password.
6. Select
7. If possible, repeat the task prior to the power failure and verify that the data were
stored.
8. If reagent was dispensed, you must perform a Weekly Wash before resuming
operation.

Additional Operating Instructions

To avoid errors or interruptions during operation, please read the following before using
the Chemistry System.

Switching Process Sequence and System Test List numbers

To avoid having quality control data reported for the wrong tests,
Sequence Numbers
affected.
To avoid incorrect reporting of results at the Review/Edit, the Daily QC windows, and the
Print Report form,
window after initial setup. If you reposition any test items at the System Test List window
after initial setup, results for samples already run could be associated with the wrong test
name.

Loading the reagent trays

CAUTION
The operator is alerted to this problem by the QC results.
Corrective action is to reload the reagents on the correct trays.

Updating the Cumulative QC

Update the QC Cumulative window before more than 200 daily control samples are run.
The QC Daily Precision Control window can manage a maximum of 200 results for each
control product (level). If an additional control sample is run, the current 200
(Px200) is overwritten by the new control result. Similarly, if multiple repetitions of a
control are requested, the system aspirates only one sample and its result replaces the
existing 200
System Overview
, then select
Re-Start
after control results have been accumulated. Patient results are also
do not modify the order of the test items
Load each reagent container (wedge) on the correct reagent tray to avoid erroneous
results when using reagent containers without a barcode. Reagent Tray 1 (RTT-1)
is the front tray, while Reagent Tray 2 (RTT-2) is the rear tray.
th
result (Px200).
.
OK
do not switch Process
at the System Test List
th
result
99