Aligning The Dpp Probe; Replenishing The Rrv (Reaction) Bath Oil Bottle - Siemens ADVIA 1800 Operator's Manual

Chemistry system
Table of Contents

Advertisement

8. Snap the joint holder (
9. Replace the splash guards and secure them back in place with the thumb screws.
10. Manually lift and rotate the probe over the probe wash cup but not within the wash
port.
11. Replace the DPP shield and secure it in place.

Aligning the DPP probe

1. Return the analyzer to the READY mode.
2. At the Menu Panel, select
3. In the Probe posi.adjust area, select Position probes for routine cleaning to move all
probes (DPP, SPP, RPP1, and RPP2) over cuvettes.
DPP should be over cuvette 1 on the DTT.
4. Ensure that the probe is perpendicular to the arm and centered over the cuvette.
If it is not, call your local technical support provider or distributor.
5. At the Operation Panel, select
water flow through the probe.
6. Verify that water is not leaking from the joint connector (

Replenishing the RRV (reaction) bath oil bottle

Materials required:
• RRV (reaction) bath oil
(REF 09323099, PN B01-4180-01)
Time: 10 minutes
Analyzer mode: STANDBY
Figure 5-33. RRV bath oil bottle
166
3
) back into the post.
, then select
Maint
User Maintenance
, then
Prime
PRIME 2
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
.
, then
to ensure proper
Execute
).
2
1
RRV (Reaction) Bath Oil
Bottl
e
2
Level Sensor Connector
3
Filter Cap
ADVIA 1800 Operator's Guide