a. Return the analyzer to the READY mode.
b. At the Operation Panel, select
c. At the PRIME Settings dialog box, select
d. Select
Cleaning and replenishing the cuvette wash bottle
Materials required:
• Deionized water
• Cuvette detergent (wash solution)
Time: 10 minutes
Analyzer mode: STANDBY
NOTE:
least once a month
Figure 5-17. Cuvette wash bottle
1. Set the analyzer to Standby.
2. Unscrew the filter cap (
tube with the filter.
3. Disconnect the cuvette wash bottle level sensor connector (
clockwise and pull it out.
142
of the number of times fields.
Execute
.
The cuvette wash bottle may be cleaned when it is refilled, but must be cleaned at
.
) at the front top of the cuvette wash bottle, then pull up the
3
WARNING
Before handling any reagents read the warnings on page 196.
CAUTION
Make a note of the bottle position on the shelf to avoid mixing up the fluid
bottles.
.
Prime
Prime 2
, then type
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
1
Cuvette Wash (Detergent)
Bottle
2
Level Sensor Connector
3
Filter Cap
ADVIA 1800 Operator's Guide
10
or more in each
), then turn it counter-
2