Emergency stop, shutdown, and recovery procedures

Stopping operation after the analyzer is started

WARNING
1. At the power panel, press the
2. Select
3. If reagents were dispensed, you must perform a weekly Wash 2 before resuming
operation.

Managing an expected power outage

If you know in advance of an upcoming power outage, do the following:
1. Turn off the workstation and analyzer power by performing the normal shutdown.
2. If you expect the power supply to be cut off for a long period of time, remove the
reagents in RTT1 and RTT2 and refrigerate them.
3. When the power returns, perform the normal startup operation.

Responding during a power failure while electric power is still off

While the electrical power is still off, do the following:
1. Turn
2. At the power panel, set the analyzer Operate/Standby switch to
When the electrical power returns, do the following:
1. Turn
2. When the Startup window opens, turn the analyzer Operate/Standby switch to
Operate
3. At the Startup window, enter a password.
4. Select
5. If possible, repeat the last task prior to the power failure, and verify that any data
generated were stored properly.
6. If reagents were dispensed, you must perform a weekly Wash 2 before resuming
operation.

To resume operations after power is restored

1. If the Startup window is open, select
operation.
2. Turn off the workstation power and turn the Operate/Standby switch on the analyzer
to
Standby
98
Make sure that all probes and mixers are free to move without obstruction and that
all analyzer covers are in place to avoid possible injury and damage to the analyzer.
.
Initialize
the workstation power switch.
OFF
the workstation power switch.
ON
.
Re-Start
, then select
.
button.
Emergency Stop
OK
.
and perform the normal shutdown
Shutdown
Standby
.
ADVIA 1800 Operator's Guide