Siemens ADVIA 1800 Operator's Manual page 170

Chemistry system
Table of Contents

Advertisement

e. Place the dryer nozzle outside the tray (1) and all the other nozzles inside the
tray.
Figure 5-35. WUD and DWUD dryer nozzle
f. Fill the tray with deionized water. Do not overflow the tray.
g. Pour enough deionized water into the tray so that the center nozzles (yellow-
labeled overflow nozzle) are in liquid.
3. At the Menu Panel, select
4. At the JEOL Maintenance window, in the Univers. sequence start area, type
Sequence field and
5. Select
• UNIVERSAL displays as the operation mode.
• One sequence takes about 65–70 seconds.
• This sequence activates the appropriate devices so all the WUD/DWUD
aspiration nozzles and overflow nozzles are pulling vacuum.
• The overflow lines (short nozzles) aspirate air after a short while as the liquid in
the container lowers. This is normal behavior.
6. When the operation mode returns to READY, lift the WUD/DWUD wash head out of
the tray and temporarily place it on top of the WUD/DWUD assembly.
7. Remove and empty the tray, then place it on paper towels under the WUD/DWUD
nozzles.
8. Place the WUD/DWUD wash head into the tray and fill the tray with 150 mL of
wash solution made up of 10% Probe Wash 3 solution (preferred) or 20% bleach
solution (alternate).
NOTE: Using more than 150 mL of wash solution may cause a leak from the vacuum
tank on the side of the instrument. If a leak occurs, secure the WUD/DWUD wash
head to the system and run a WASH3.
170
Maint
1
in the Number of times field.
Start
, then select
Yes
at the confirmation window to start the procedure.
., then select
JEOL Maintenance
ADVIA 1800 Operator's Guide
.
in the
1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents