Siemens ADVIA 1800 Operator's Manual page 152

Chemistry system
Table of Contents

Advertisement

5. Remove the cap (
detergent solution into the dummy electrode.
6. Replace and tighten the cap.
7. At the ISE Operation window, select
The message "Line Wash 1 Running" and the approximate amount of time remaining
displays. You cannot stop this operation. If you must stop, press the
button on the analyzer display panel. This step takes approximately 17 minutes to
complete.
This step takes about 17 minutes.
8. When step 1 of the wash completes, replace the dummy electrode with the original
Na, K, Cl, and Ref electrodes.
9. At the ISE Operation window, in the ISE line wash area, select
The wash starts and buffer prime is performed 10 times. The message "Line Wash 2
Running" and the approximate amount of time remaining displays.
10. Verify no leaks or bubbles exist and that the buffer is going to the waste during the
priming cycle.
11. At the ISE Operation window, next to the Initialize area, select
12. Perform calibration and run controls.
13. Perform an additional 10 buffer primes, then calibrate, and run controls.
152
1
) from the dummy electrode and pour approximately 5 mL of ISE
CAUTION
Be sure to tighten the cap on the dummy electrode. If the cap is loose or
defective, the ISE detergent solution may leak into the module and cause
damage.
.
Execute (STEP-1)
ADVIA 1800 Operator's Guide
SYSTEM STOP
Execute (STEP-2)
.
Execute