Testing A Slotless
contact
Slotless Contact Repair
After you install a new inner contact, you must test it using the
tools provided in this repair kit. This tool was designed to help nd
improperly installed inner contacts.
1. Using a foam swab and alcohol, clean the testing tool.
2. Under magnication, carefully install the testing tool in the
slotless contact assembly (see Figure 2-16).
3. Repeat step 2, two more times.
4. Inspect the inner contact. If the ngers are damaged, remove and
replace the inner contact.
5. Insert the testing tool.
6. With the testing tool inserted, turn the device upside down so
that the testing tool hangs by the grip of the slotless contact.
7. Hook the testing weight to the testing tool. If the contact has
the proper minimum retention force, it does not lose its grip.
8. Remove the testing weight.
9. Remove the testing tool.
10. Visually inspect the slotless contact.
11. If you see any dirt or metal particles, clean the contact using the
alcohol and foam swabs provided in this kit.
12. The clean, precision slotless connector is ready for use.
Figure 2-16. Inserting the Testing Tool
Repairing A Slotless Contact
2-11