Sirona CEREC MC XL Operating Instructions Manual page 20

Dental milling machine
Table of Contents

Advertisement

5 Installation and startup
Initial startup
20
LAN switch, connection example with 2 milling units
1.
Connect the LAN port (A) of the MC XL milling unit with the LAN switch
using the included 10m LAN network cable.
2.
Connect the access point with the LAN switch using the additional 10m
LAN network cable.
Now, all MC XL milling units connected to the LAN switch can be operated
via WLAN.
5.4.5.6
Connecting to the PC via the wireless H&W interface
(optional)
✔ The CEREC AC is equipped with a wireless module HW 8614/F2.
Installation kit: 62 79 694
Retrofit kit: 62 79 702
1.
Connect the wireless module HW 8614/F2 to the LAN port of the MC XL,
using the LAN crossover cable (1 m).
2.
Connect the plug-in power supply included in the scope of delivery with
the wireless module HW 8614/F2 and plug it into the power supply.
3.
If applicable, fasten the wireless module in the selected operating
position, using the pre-installed Velcro strap.
Make sure that the rod antenna is in vertical position.
4.
Pair the wireless module of the milling unit as described in the installation
instructions included with the wireless module (Order Number 62 80
064).
You can pair multiple milling units with a CEREC AC. If more than 2 milling
units are operated at once, the limited data bandwidth may cause milling time
delays.