Preparing optical scanning (only with integrated scanner)
34
Fastening the model
Position of the model
Fastening the model
A
Allen screw
✔ The preparation margin is not visible from all sides.
➢ Fasten the model in the correct position with the Allen screw.
Valid scan range of the model:
The lowest area must not lie more than 2 mm below the center line of the
model holder.
The highest area must not lie more than 15 mm above the center line of
the model holder.
6.5.3

Preparing for scanning of the veneer model

✔ You have made a model of the clinical situation in the usual manner.
➢ Glue this model onto the model holder. The prepared labial surface must
point toward the slot on the model holder.
The holder is automatically turned so that the scanning process can be
performed from the labial direction.
Veneer model
A
Slot on model holder