5 Installation and startup
Initial startup
22
9.
Switch the pump on (press the
10. Fill the water tank up again until the filter insert is completely immersed
(up to the bottom edge of the cap thread).
5.4.7

Switching the unit ON and OFF

CAUTION
Note on condensate
Do not put the unit into operation at low temperatures!
If you move the unit to the operating site from a cold environment,
condensation may form and result in a short circuit.
The milling unit contains grease depots for lubricating components which
can cause error messages at low temperatures.
✔ Install the unit at room temperature.
➢ Wait until the unit has reached room temperature and is absolutely dry
(for at least one hour)
The unit is dry and can be put into operation.
NOTICE
Line voltage
Do not adjust the line voltage
The unit automatically adjusts to the line voltage.
Switching the unit on
✔ The milling unit is connected to the power supply.
1.
The main switch on the rear side of the unit is set to position I (ON).
2.
Press the ON/OFF button on the front panel.
The unit switches on and the display lights up.
192. 168. 230. 120
Power-up display
When the milling unit is switched on, the display shows a picture of the milling
unit trying to contact the PC.
You can start or stop the water pump by pressing the
enables you to drain the water circuit without connecting to the PC (e.g. prior
to transport) or fill the water circuit during startup.
You can call up the IP address by pressing the
configure the milling unit in the network with this address.
Sirona Dental Systems GmbH
Operating Instructions CEREC MC XL
"Pump"
button), to fill the water circuit.
?
Config
Pump
C
D
"Pump"
button (C). This
"Config"
button (D). You can