7 Maintenance
Changing the water
38
7.1
Changing the water
7.1.1
General information
CAUTION
Damage to the pump and milling drives!
An excessively high ceramic content in the cooling water will damage the
pump and milling drives.
Change the water regularly!
When it is time for the water to be changed, a message window appears on
your monitor to remind you that it is time to change the water.
Preventing odors
NOTICE
Odors!
All milling additives contain a biologically degradable preservative. Despite
this, however, odors may still develop under unfavorable conditions.
Observe the following:
●
Change the water at least once a week.
●
With ambient temperatures above 25°C, change the water every 2 to 3
days to prevent foul odors.
●
Drain the tank if you do not intend to operate the unit for more than one
week.
●
Clean the tank if the odors recur.
●
Add DENTATEC milling additive and fill the tank up to the brim with water.
Let it stand for at least 24 hours and then rinse it out thoroughly with water
once again.
CAUTION
Damage to surfaces!
DENTATEC milling additive etches plastic surfaces in undiluted form and
can cause discoloration.
➢ Do not place DENTATEC on the unit.
➢ Do not spill DENTATEC.
CAUTION
Permissible milling additive
Use only DENTATEC as a milling additive.
Sirona Dental Systems GmbH
Operating Instructions CEREC MC XL
61 46 885 D 3439
D 3439.201.02.11.02 10.2009