In all of these cases, you can deactivate sets 1 and 2 separately in the
Configuration"
dialog box. A deactivated set will simply be ignored during
milling, calibration etc.
6.2.2
Calibrating the scanner (only with integrated
scanner)
NOTICE
inLab + CEREC scanner calibration
Unit calibrated ex works
The unit is calibrated at the factory. No additional calibration is required
during initial startup. Proceed as described below when performing a
subsequent calibration.
✔ Remove the calibration body from the drawer of the unit.
1.
In the menu line, select the command
"Scanner"
.
2.
If several milling units are connected, a dialog box will appear. Select the
milling unit you prefer and confirm your choice with
The milling unit then moves into position to insert the calibration body.
A dialog box prompts you to insert the calibration body.
3.
Press the catch of the milling chamber door and open the door.
NOTICE
Do not remove the milling instruments
The milling instruments can remain in the motor mounts during the
operation.
6 Operation
Calibrating the unit
"Device
"Settings"
/
"Calibration"
/
"OK"
.
27