7.6

Changing the filter

NOTICE
Change the filter regularly!
Clean the filter regularly and change it immediately when damaged.
Otherwise, change it every 3 months.
If a message appears stating that the water pressure is too low, you must
clean the filter or, if it is damaged, replace it immediately.
CAUTION
Filter
Use only filters approved by Sirona!
A
CEREC water tank
Water tank
A
Filter insert
B
Tank cap
✔ The tank is drained, see "Removing water from the unit" [ 46].
1.
Pull out the water tank at the front of the unit.
2.
Unscrew the cover on the side and take it out of the tank along with the
filter insert.
3.
Rinse the water tank.
4.
Insert a new filter with cover into the tank and screw it tight.
5.
Fill the tank, see "Changing the water" [ 38].
Filter insert:
Order No. 61 29 519
B
D
C
C
Tank
D
Tank drain