Calibration Settings; Format - Keysight N9923A User Manual

Fieldfox rf network analyzers
Hide thumbs Also See for N9923A:
Table of Contents

Advertisement

NA (Network Analyzer) Mode
NA Mode Settings

Calibration Settings

NA Mode has a page specifically for making settings that pertain to Port
Extensions, including Velocity Factor and Media Type.
Learn more in
Learn more about Media Type in
How to view and change Calibration Settings
— Press Meas Setup 4.
— Then
— Make these settings in the same manner as Quick Settings in the previous

Format

Select the display format in which to present measurement results. This setting
can be changed at any time without affecting calibration accuracy.
Learn more about Display Formats in the FieldFox Supplemental Online
Help:
http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htm
The marker on-screen readout can be changed to formats other than the
display format. Learn more in
How to select a NA Format
— Press Measure 1.
— Then
— Then choose from the following:
46
—Full (Default setting) Only the Settings table is shown on the screen.
The trace window is temporarily not shown.
—Left The Settings table is shown to the left of the trace window.
—Bottom The Settings table is shown below the trace window.
When finished changing ALL settings, press
"Port Extensions" on page

Calibration Settings

section.
Format [current
—Log Magnitude
Displays magnitude in dB
—Linear
Displays positive values only. Y-axis: Unitless (U) for ratioed
measurements; Watts (W) for unratioed measurements.
—VSWR
Used mainly for S11 and S22. Displays unitless reflection data.
—Phase
Displays phase in degrees. The trace 'wraps' every 360 degrees,
from +180 to –180, for easy scaling.
Done
54.
"Waveguide Calibrations" on page
"Marker Format" on page
setting].
Keysight N9923-90001 User's Guide
to save your settings.
117.
178.