Interference Rejection - Keysight N9923A User Manual

Fieldfox rf network analyzers
Hide thumbs Also See for N9923A:
Table of Contents

Advertisement

CAT (Cable and Antenna Test) Mode
CAT Mode Settings

Interference Rejection

Use this setting when you suspect that other signals in the area are interfering
with a measurement. Interference may look like a spike or lack of stability in the
measurement trace. While monitoring a measurement at a specific frequency,
toggle this setting between ON and OFF. If the measurement result decreases
while ON, then there is an interfering signal in the area. Continue to make
measurements with Interference Rejection ON. However, this will slow the
measurement speed.
Once enabled, up to SIX sweeps may be required before the interfering signal
is neutralized.
This setting can be changed at any time without affecting calibration accuracy.
How to set Interference Rejection
— Press Meas Setup 4.
— Then
— Then choose from the following:
Keysight N9923-90001 User's Guide
Interference Rejection [current
—Off
No interference rejection and fastest possible sweep speed
—Minimum
The lowest level of Interference rejection.
—Med ium
The medium level of Interference rejection.
—Maximum
The highest level of Interference rejection.
setting].
85

Advertisement

Table of Contents
loading

Table of Contents