Charging The Battery; Internal Charging With The Ac/Dc Adapter - Keysight N9923A User Manual

Fieldfox rf network analyzers
Hide thumbs Also See for N9923A:
Table of Contents

Advertisement

Working with the Lithium-Ion Battery

Charging the Battery

The battery in your FieldFox will have either 4 or 5 LED segments.

Charging the Battery

When you receive your FieldFox, the lithium-ion battery is partially charged to
approximately 30% to 40% to preserve battery life. Allow four hours to fully
charge the battery internally, or three hours by using the external battery
charger (N9910X–872). A fully charged battery will power your FieldFox for
about four hours. It is NOT necessary for you to charge the battery.
The FieldFox circuitry ALWAYS prevents the battery from discharging to a level
that is damaging. However, if the battery is completely discharged, the
FieldFox may not turn ON. If this occurs, the battery can still be charged either
internally or externally.
Because a lithium-ion battery has no memory effect, you can charge a battery
at any time regardless of the current charge status. However, this could impair
the accuracy of its internal charge-remaining indicator. When the Max Error is
greater than 10%, the battery should be reconditioned. Learn more in
"Reconditioning Battery" on page

Internal Charging with the AC/DC Adapter

Read the safety information in
Internal charging time for a fully depleted battery is approximately 4 hours with
the FieldFox either ON or OFF.
Keysight N9923-90001 User's Guide
5-segment
Charge
Gauge
Remaining
5 segments
≥80%
4 segments
60% to 79%
3 segments
40% to 59%
2 segments
20% to 39%
1 segment
< 20%
4-segment Gauge
4 segments
3 segments
2 segments
1 segment
244.
"For the AC/DC Adapter" on page 250
Charge Remaining
≥75%
50%,to 74%
25% to 49%
< 25%
.