CAT (Cable and Antenna Test) Mode
CAT Mode Settings

Output Power

Set the power level out of the FieldFox to High, Low, or manually set power
level to a value between High and Low.
Generally, the high power setting is used when measuring passive, high-loss
devices to place the signal farther from the noise floor. However, for devices
that are sensitive to high power levels such as amplifiers, use the Low power
setting.
For best measurement accuracy, use the Manual power setting at -15 dBm.
After calibration, the power level can be decreased for amplifiers, or increased
for higher dynamic range.
Power Level settings in this mode will NOT change Power Level settings in
other modes. To help prevent damage to your DUT, use caution when
changing modes with your DUT connected to the FieldFox test ports.
How to set Output Power
— Press Meas Setup 4.
— Then
— Then
Keysight N9923-90001 User's Guide
Power

Output Power

—High (Default setting) Sets output power to the maximum achievable
power at all displayed frequencies. Output power is NOT FLAT across
the displayed FieldFox frequency span. Please see
"Specifications/Data Sheet" on page 263
—Low Sets output power to approximately –45 dBm, FLAT across the
displayed FieldFox frequency span.
—Man Set output power to an arbitrary value, FLAT across the displayed
FieldFox frequency span. If flattened power can NOT be achieved, a
warning message and beep occurs. To achieve a flattened output
power, reduce the power level or stop frequency.
—Then press
Nominal Power
—Then enter a value using the numeric keypad, the
rotary knob.
for expected power levels.